An ultrasound simulation method for carotid arteries with a wall structure of three membranes

在线阅读 下载PDF 导出详情
摘要 Ultrasoundsimulationforcarotidarteriesishelpfultotheperformanceassessmentsofvesselwalldetectionandsignalprocessingmethodsbyusingultrasoundtechniques.Anultrasoundsimulationmethodofcarotidarterywallwithathree-membranestructureisproposedinpresentstudy.Accordingtotheultrasoundspeckledistributionsvaryingwiththeshapesanddensitiesofscattererdistributions,aswellasthestatisticresultsoftheclinicalimages,theparametersofdistributions,densitiesandintensitiesofscatterersfordifferentkindsoftissuesinthecarotidarteryphantomsaredetermined.EachregionisacousticallycharacterizedusingFIELDIIsoftwaretoproducetheradiofrequencyechosignals,fromwhichultrasoundimagesarederived.Theresultsbasedon30simulationsshowthattheechodistributionsoftheintimae,mediae,adventitiasandbloodareconsistentwiththeclinicalones.Moreover,comparedwiththeresultsfromthecentralfrequencyof8MHz,themeanmeasurementsforthicknessesoftheintima,mediaandadventitiamembranes,aswellasthelumendiameterfromthesimulationimagesbasedon12MHzarethesameasthepresetones,andthemaximumrelativeerrorsarethe4.01%,1.25%,0.04%and0.15%,respectively.Thesimulationunderthisconditionismorerealistic.
机构地区 不详
出处 《声学学报:英文版》 2015年3期
出版日期 2015年03月13日(中国期刊网平台首次上网日期,不代表论文的发表时间)
  • 相关文献