简介:目的:用Tono-Pen眼压计测量人眼角膜中央、旁中央、角巩缘部位眼压值,比较不同部位值的差异及相关性。方法:正常角膜用Tono-Pen眼压计依次测量角膜中心、旁中心、角膜缘眼压值,方差分析不同部位眼压的差异,分析结果的相关性。结果:Tono-Pen眼压计检测得出的眼压平均值角膜中央为16.28±2.73mmHg,旁中央为16.33±2.69mmHg,角巩缘为16.58±2.58mmHg.角膜中心与旁中心的相关系数r=0.966,P=0.000,角膜中心与角膜缘的相关系数为r=0.897.P=0.000,角膜旁中心与角膜缘的相关系数为r=0.910,P=0.000。不同部位眼压值差异没有显著性(F=0.093,P=0.913〉0.05)。结论:不同部位眼压值密切相关,差异没有统计学意义。Tono-Pen眼压计测量不同部位眼压均可取得较为一致的结果。
简介:目的:通过与非接触式眼压计(non-contacttonometer,NCT)的比较,评价Diaton眼压计(DT)在准分子激光原位角膜磨镶术(laserinsitukeratomileusis,LASIK)前后眼压测量中的应用价值。方法:在LASIK术前及术后1mo,分别使用NCT和DT测量患者眼压,使用SPSS13.0统计软件对所得结果进行统计学分析。结果:术后1mo时,NCT眼压测量值较术前下降(5.76±2.31)mmHg,DT眼压测量值则无显著差异(t=0.70,P=0.487);NCT与DT读数测量值差值与术中切削量呈正相关,与术后中央角膜厚度呈负相关。结论:Diaton测量眼压不受角膜厚度的影响,因此更适用于LASIK术后患者的眼压测量。
简介:AIM:Tofindasimplemathematicalcorrelationbetweenthelensbasecurve(BC)andkeratometryfindings(krf).METHODS:Thisretrospectivestudyincluded400keratoconiceyes(350patients)previouslyfitwithrigidcontactlensesatanacademiceyecenteroverafiveyearperiod.Thepatientswereclassifiedintofivegroupsbasedonthekeratometryfindings(krf<7,krf:7-8,krf>8,krf-krs(differencebetweentwokeratometry;flatandsteep)=0.3-0.6,krf-krs>0.6mmasgroups1to5,respectively.MultivariatelinearregressionandMunro’scorrelationcoefficientwereemployedtodefertheformulas.RESULTS:Alinearcorrelationcouldbefoundinallgroupsexceptforpatientsingroup3.Forgroup1,BC=0.211×krf+5.904.Forgroup2,BC=0.456×krf+4.160.Forgroup4,BC=0.321×krf+5.219.Forgroup5,BC=0.337×krf+5.090.CONCLUSION:ThedevelopmentofnewformulasforRGPfittingenablesophthalmologiststoworkwithconfidenceandpreventsunnecessaryandfrequentlenstrials.Thecustomarylensfittingmethodsareneededtobereplacedbynewformulas,whichhelptosavetimeandcosts.
简介:目的:探讨裂隙灯眼前段处理系统在眼科临床工作中的各种实际应用状况及操作技巧。方法:应用配置佳能PowerShotA720IS型数码照相机(1200万像素)SLM型裂隙灯显微镜检查眼部病变情况,并在裂隙灯下根据不同的病变位置,在不同色彩、角度下进行照相(放大倍率×10;×16;×20)。结果:采集不同种类疾病具有代表性照片:眼睑及结膜肿物、结膜裂伤、角膜炎、角膜异物、翼状胬肉、前房积血、前房角异物等如图示。结论:裂隙灯眼前段处理系统的应用为临床医疗文献提供直接定性依据,给患者了解自身病情带来便利,照片直观、经济,在眼科领域的临床应用具有广阔的前景。
简介:目的评价海德堡视网膜断层扫描仪黄斑区视网膜分析系统(macularedemamodule,MEM)的三个技术参数,了解其变异性及可重复性来判断其临床应用价值。方法对78例(78只眼)正常健康志愿者应用海德堡视网膜断层扫描仪黄斑区视网膜分析系统进行检查。分析扫描深度值(Z)、视网膜信号宽度值(W)、水肿指数值(E)三参数的均值及变异系数,并分析黄斑反射图、地形图、信号宽度图、水肿指数图的图像特征。结果黄斑中心0.5mm范围内视网膜扫描深度均值(Z)为1.4430848,变异系数55.6%。W值为795.27±193.53,变异系数17.6%。E值为1.13±0.35,变异系数27.8%。不同个体三参数差异有高度显著性(P<0.01),个体变异较大。不同时间段三参数测量差异无显著性(P>0.05)。结论MEM可对黄斑区视网膜厚度进行量化分析,个体变异较大,但重复性好,HRT-Ⅱ检查适合黄斑疾病的个体随诊观察。
简介:目的探讨数字化多媒体系统矫治训练对大龄儿童弱视的治疗效果。方法选取2013年1月至2016年1月我院门诊收治的大龄弱视患儿86例(128眼),作为研究对象。根据患儿治疗方案的差异将其分为观察组与对照组,对照组45例患儿接受常规综合治疗措施,观察组41例患儿接受数字化多媒体系统治疗,比较两组患儿治疗后双眼视功能变化及疗效。结果观察组患儿治疗后总有效率明显高于对照组,并且观察组患者立体视改善率明显高于对照组,P〈0.05,差异具有统计学意义。结论对大龄弱视患儿采取数字化多媒体系统治疗可有效改善患儿弱视症状,提高患儿双眼视功能,缩短患儿视功能障碍治疗时间,具有临床应用及推广价值。
简介:AIM:Anaerobicbacteriacancauseocularinfections.WetestedtheOxyPlateTMAnaerobicSystem(OXY)toisolatepertinentanaerobicbacteriathatcancauseoculardisease.METHODS:OXY,whichdoesnotrequiredirectanaerobicconditions(i.e.bags,jars),wascomparedtoconventionalisolationofincubatingculturemediainanaerobicbags.Standardcoloniescountswereperformedonanaerobicocularbacterialisolatesunderaerobicandanaerobicconditions(anaerobicbags)usingagarmedia:1)OXY(aerobiconly),2)5%sheepblood(SB),3)Chocolate,and4)Schaedler.Thebacteriatestedwerede-identifiedocularisolatesculturedfromendophthalmitisanddacryocystitisthatinclude10Propionibacteriumacnesand3Actinomycesspecies.Thecolonycountsforeachbacteriaisolate,oneachculturingcondition,wererankedfromlargesttosmallest,andnon-parametricallycomparedtodeterminethebestculturingcondition.RESULTS:Allanaerobicconditionswerepositiveforalloftheanaerobicisolates.SBandSchaedler’sagarunderaerobicconditionsdidnotsupportthegrowthofanaerobicbacteria.SparsegrowthwasnotedonchocolateagarwithPropionibacteriumacnes.Asananaerobicsystem,SBinananaerobicbagisolatedhighercolonycountsthanOXY(P=0.0028)andchocolateagar(P=0.0028).CONCLUSION:AlthoughOXYdidnottesttobemoreefficientthanotheranaerobicsystems,itappearstobeareasonablealternativeforisolatinganaerobicbacteriafromocularsites.Theuseofanagarmediuminaspeciallydesignedplate,withouttherequirementofananaerobicbag,renderedOXYasanadvantageoverotheranaerobicsystems.
简介:AIM:Tomeasurecentralcornealthickness(CCT)andpre-cornealtearfilmthicknessusingtheGalileidualScheimpfluganalyzer(GSA)inNewZealandWhiterabbits.METHODS:TennormalNewZealandWhiterabbits(20eyes)wereincludedinthisstudy.Withtheassistanceof0.1%fluorescein,thepre-cornealtearfilmcanbewellvisualized.BotheyesofeachrabbitwerescannedoncewiththeGSApre-andpost-instillationof1μL0.1%fluorescein.ThedifferencebetweenthetwomeasurementsofCCT(4-mmdiameter)wasrecordedasthepachymetricvaluesofthecentraltearfilm.RESULTS:TheCCTofpre-andpost-instillationwas388.8±9.5μmand407.0±10.5μm,respectively.Afterapairedt-testanalysis,thecentralpre-cornealtearfilmthicknessof4mmdiameterwas18.2±5.31μmwitha95%confidenceintervalof(15.7,20.6)μm(P<0.001).CONCLUSION:GSAcanbeusedtomeasureCCTandanalyzecentraltearfilmthicknessofrabbitswiththehelpoffluorescein.
简介:AIM:Toevaluatetheaccuracyofsphericalequivalent(SE)estimatesofadouble-passsystemandtocompareitwithretinoscopy,subjectiverefractionandatablemountedautorefractor.METHODS:Non-cycloplegicrefractionwasperformedon125eyesof65healthyadults(age23.5±3.0years)fromOctober2010toJanuary2011usingretinoscopy,subjectiverefraction,autorefraction(AutokeratorefractometerTOPCONKR-8100,Japan)andadoublepasssystem(OpticalQualityAnalysisSystem,OQAS,VisiometricsS.L.,Spain).Nineconsecutivemeasurementswiththedouble-passsystemwereperformedonasubgroupof22eyestoassessrepeatability.ToevaluatethetruenessoftheOQASinstrument,theSElaboratorybiasbetweenthedoublepasssystemandtheothertechniqueswascalculated.RESULTS:TheSEmeancoefficientofrepeatabilityobtainedwas0.22D.SignificantcorrelationscouldbeestablishedbetweentheOQASandtheSEobtainedwithretinoscopy(r=0.956,P<0.001),subjectiverefraction(r=0.955,P<0.001)andautorefraction(r=0.957,P<0.001).ThedifferencesinSEbetweenthedouble-passsystemandtheothertechniquesweresignificant(P<0.001),butlackedclinicalrelevanceexceptforretinoscopy;Retinoscopygavemorehyperopicvaluesthanthedouble-passsystem-0.51±0.50Daswellasthesubjectiverefraction-0.23±0.50D;Moremyopicvalueswereachievedbymeansofautorefraction0.24±0.49D.CONCLUSION:Thedouble-passsystemprovidesaccurateandreliableestimatesoftheSEthatcanbeusedforclinicalstudies.Thistechniquecandeterminethecorrectfocuspositiontoassesstheocularopticalquality.However,ithasarelativelysmallmeasuringrangeincomparisonwithautorefractors(-8.00to+5.00D),andrequirespriorinformationontherefractivestateofthepatient.